A dirty window only ruins the reading if you are measuring the speed of the oxygen passing through it. The three electrode design stopped measuring speed and started measuring balance. Unless the gunk is a total airtight seal (which is rare on the scale of an oxygen molecule), the sensor will eventually reach the right answer, whereas the old version would fail.
Coeur|1 month ago
gorjusborg|1 month ago
The big gain comes from a change in how you interpret the presence of electrons.
The older approach converted oxygen to electrical current, the magnitude of current flow relating to magnitude of oxygen depletion. The assumption built into that approach is that low oxygen depletion levels meant low oxygen levels, but that wasn't the only potential cause, because it ignored variation in the permiability of the membrane.
The newer approach equates current flow to oxygen concentration, as the system doesn't deplete the concentration any longer. The permiability of the membrane in this setup only contributes to a longer initial delay as the inner chamber comes to equilibrium with the surrounding concentration.